Test Date: 2015-08-24 17:49
Analysis date: 2016-05-26 02:29
Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center Center of Peak 49.97 ± 0.07 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.73
Rate Rate 230.60 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center Center of Peak 100.2 ± 0.07 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.87
Rate Rate 390.40 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center Center of Peak 127.23 ± 0.1 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 4.78
Rate Rate 224.00 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center Center of Peak 144.53 ± 0.12 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 5.08
Rate Rate 146.30 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope Slope 48.772 ± 0.059 nElectrons/VCal
Offset Offset -36.32 ± 5.797 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate Rate 230.60 kHz/cm²
NTrigs N Trig 5920163 Trigger
NHits N Hits 21293 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate Rate 390.40 kHz/cm²
NTrigs N Trig 5919960 Trigger
NHits N Hits 36054 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate Rate 224.00 kHz/cm²
NTrigs N Trig 5920333 Trigger
NHits N Hits 20687 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate Rate 146.30 kHz/cm²
NTrigs N Trig 5919857 Trigger
NHits N Hits 13513 Hits
MoReWeb-v1.0.0-5-g82d9ff6 on branch psi46master