Test Date: 2015-05-21 18:05
Chip 12
X-ray Calibration - Target Zn Method Spectrum Chip C12
Center Center of Peak 54.57 ± 0.07 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 19.48
Rate Rate 340.40 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C12
Center Center of Peak 106.76 ± 0.06 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 11.7
Rate Rate 534.20 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C12
Center Center of Peak 136.43 ± 0.1 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 7.14
Rate Rate 287.70 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C12
Center Center of Peak 154.64 ± 0.11 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 7.41
Rate Rate 181.30 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C12
Slope Slope 44.879 ± 0.108 nElectrons/VCal
Offset Offset 60.206 ± 13.323 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C12
Rate Rate 340.40 kHz/cm²
NTrigs N Trig 5910555 Trigger
NHits N Hits 31385 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C12
Rate Rate 534.20 kHz/cm²
NTrigs N Trig 5910983 Trigger
NHits N Hits 49261 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C12
Rate Rate 287.70 kHz/cm²
NTrigs N Trig 5941040 Trigger
NHits N Hits 26666 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C12
Rate Rate 181.30 kHz/cm²
NTrigs N Trig 5909762 Trigger
NHits N Hits 16712 Hits