Test Date: 2015-05-21 18:05
Chip 3
X-ray Calibration - Target Zn Method Spectrum Chip C3
Center Center of Peak 54.86 ± 0.08 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 51.05
Rate Rate 336.10 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C3
Center Center of Peak 105.7 ± 0.06 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 22.31
Rate Rate 530.80 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C3
Center Center of Peak 134.91 ± 0.1 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 11.6
Rate Rate 284.80 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C3
Center Center of Peak 153.36 ± 0.11 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 10.39
Rate Rate 182.10 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C3
Slope Slope 45.206 ± 0.109 nElectrons/VCal
Offset Offset 74.693 ± 13.351 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C3
Rate Rate 336.10 kHz/cm²
NTrigs N Trig 5910555 Trigger
NHits N Hits 30988 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C3
Rate Rate 530.80 kHz/cm²
NTrigs N Trig 5910983 Trigger
NHits N Hits 48946 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C3
Rate Rate 284.80 kHz/cm²
NTrigs N Trig 5941040 Trigger
NHits N Hits 26398 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C3
Rate Rate 182.10 kHz/cm²
NTrigs N Trig 5909762 Trigger
NHits N Hits 16790 Hits