Test Date: 2015-09-30 20:12
Analysis date: 2015-10-26 19:40
Chip 15
X-ray Calibration - Target Zn Method Spectrum Chip C15
Center Center of Peak 45.42 ± 0.13 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.31
Rate Rate 92.70 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C15
Center Center of Peak 90.98 ± 0.18 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 1.64
Rate Rate 90.80 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C15
Center Center of Peak 116.21 ± 0.19 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 3.02
Rate Rate 39.50 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C15
Center Center of Peak 133.48 ± 0.42 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 3.88
Rate Rate 19.20 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C15
Slope Slope 51.266 ± 0.431 nElectrons/VCal
Offset Offset 193.221 ± 45.986 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C15
Rate Rate 92.70 kHz/cm²
NTrigs N Trig 9905134 Trigger
NHits N Hits 14322 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C15
Rate Rate 90.80 kHz/cm²
NTrigs N Trig 9903197 Trigger
NHits N Hits 14021 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C15
Rate Rate 39.50 kHz/cm²
NTrigs N Trig 39623229 Trigger
NHits N Hits 24441 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C15
Rate Rate 19.20 kHz/cm²
NTrigs N Trig 59422274 Trigger
NHits N Hits 17807 Hits
MoReWeb-v0.6.7-11-ga66536c on branch psi46master