Test Date: 2016-02-08 11:28
Analysis date: 2016-05-26 09:02
Chip 8
X-ray Calibration - Target Zn Method Spectrum Chip C8
Center |
Center of Peak |
54.48 ± 0.07 |
Vcal |
TargetEnergy |
Energy of target Zn |
8638.91 |
eV |
TargetNElectrons |
Energy of target Zn |
2399.7 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
1.61 |
|
Rate |
Rate |
315.30 |
kHz/cm² |
X-ray Calibration - Target Mo Method Spectrum Chip C8
Center |
Center of Peak |
108.51 ± 0.07 |
Vcal |
TargetEnergy |
Energy of target Mo |
17479.37 |
eV |
TargetNElectrons |
Energy of target Mo |
4855.38 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
2.4 |
|
Rate |
Rate |
467.00 |
kHz/cm² |
X-ray Calibration - Target Ag Method Spectrum Chip C8
Center |
Center of Peak |
136.88 ± 0.1 |
Vcal |
TargetEnergy |
Energy of target Ag |
22162.92 |
eV |
TargetNElectrons |
Energy of target Ag |
6156.37 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
4.4 |
|
Rate |
Rate |
246.80 |
kHz/cm² |
X-ray Calibration - Target Sn Method Spectrum Chip C8
Center |
Center of Peak |
154.79 ± 0.13 |
Vcal |
TargetEnergy |
Energy of target Sn |
25271.36 |
eV |
TargetNElectrons |
Energy of target Sn |
7019.82 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
3.12 |
|
Rate |
Rate |
157.00 |
kHz/cm² |
X-ray Calibration - Vcal Calibration Method Spectrum Chip C8
Slope |
Slope |
45.545 ± 0.062 |
nElectrons/VCal |
Offset |
Offset |
-82.911 ± 6.048 |
nElectrons |
X-ray Hit Map - Target Zn Method Spectrum Chip C8
Rate |
Rate |
315.30 |
kHz/cm² |
NTrigs |
N Trig |
5915565 |
Trigger |
NHits |
N Hits |
29097 |
Hits |
X-ray Hit Map - Target Mo Method Spectrum Chip C8
Rate |
Rate |
467.00 |
kHz/cm² |
NTrigs |
N Trig |
5915687 |
Trigger |
NHits |
N Hits |
43096 |
Hits |
X-ray Hit Map - Target Ag Method Spectrum Chip C8
Rate |
Rate |
246.80 |
kHz/cm² |
NTrigs |
N Trig |
5916091 |
Trigger |
NHits |
N Hits |
22773 |
Hits |
X-ray Hit Map - Target Sn Method Spectrum Chip C8
Rate |
Rate |
157.00 |
kHz/cm² |
NTrigs |
N Trig |
5915804 |
Trigger |
NHits |
N Hits |
14493 |
Hits |
MoReWeb-v1.0.0-5-g82d9ff6 on branch psi46master