Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center Center of Peak 51.81 ± 0.07 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.61
Rate Rate 241.10 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center Center of Peak 103.11 ± 0.08 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 3.33
Rate Rate 393.40 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center Center of Peak 130.03 ± 0.1 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 5.05
Rate Rate 232.00 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center Center of Peak 147.95 ± 0.14 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 4.33
Rate Rate 151.90 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope Slope 48.288 ± 0.157 nElectrons/VCal
Offset Offset -123.373 ± 18.956 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate Rate 241.10 kHz/cm²
NTrigs N Trig 5917211 Trigger
NHits N Hits 22252 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate Rate 393.40 kHz/cm²
NTrigs N Trig 5916911 Trigger
NHits N Hits 36314 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate Rate 232.00 kHz/cm²
NTrigs N Trig 5916637 Trigger
NHits N Hits 21413 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate Rate 151.90 kHz/cm²
NTrigs N Trig 5916877 Trigger
NHits N Hits 14021 Hits