Test Date: 2015-05-26 14:05
Chip 11
X-ray Calibration - Target Zn Method Spectrum Chip C11
Center Center of Peak 45.89 ± 0.09 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 6.2
Rate Rate 354.00 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C11
Center Center of Peak 100.12 ± 0.08 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 4.15
Rate Rate 554.10 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C11
Center Center of Peak 127.95 ± 0.15 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 6.9
Rate Rate 294.90 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C11
Center Center of Peak 146.6 ± 0.16 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 16.67
Rate Rate 225.50 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C11
Slope Slope 46.61 ± 0.165 nElectrons/VCal
Offset Offset 189.226 ± 18.845 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C11
Rate Rate 354.00 kHz/cm²
NTrigs N Trig 5911837 Trigger
NHits N Hits 32647 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C11
Rate Rate 554.10 kHz/cm²
NTrigs N Trig 5911915 Trigger
NHits N Hits 51104 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C11
Rate Rate 294.90 kHz/cm²
NTrigs N Trig 5911947 Trigger
NHits N Hits 27201 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C11
Rate Rate 225.50 kHz/cm²
NTrigs N Trig 5911790 Trigger
NHits N Hits 20801 Hits