Test Date: 2015-04-08 11:04
Chip 10
X-ray Calibration - Target Cu Method Spectrum Chip C10
Center Center of Peak 50.46 ± 0.11 Vcal
TargetEnergy Energy of target Cu 8047.82 eV
TargetNElectrons Energy of target Cu 2235.51 nElectrons
Chi2PerNDF Chi^2 per NDF 7.02
Rate Rate 361.90 kHz/cm²
X-ray Calibration - Target Zn Method Spectrum Chip C10
Center Center of Peak 53.23 ± 0.11 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 6.31
Rate Rate 407.20 kHz/cm²
X-ray Calibration - Target Br Method Spectrum Chip C10
Center Center of Peak 71.36 ± 0.13 Vcal
TargetEnergy Energy of target Br 11924.36 eV
TargetNElectrons Energy of target Br 3312.32 nElectrons
Chi2PerNDF Chi^2 per NDF 5.45
Rate Rate 618.90 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C10
Center Center of Peak 106.75 ± 0.1 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 4.11
Rate Rate 573.50 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C10
Center Center of Peak 135.98 ± 0.15 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 3.65
Rate Rate 293.40 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C10
Center Center of Peak 155.13 ± 0.15 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 3.59
Rate Rate 185.70 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C10
Slope Slope 44.697 ± 0.159 nElectrons/VCal
Offset Offset 83.113 ± 20.147 nElectrons
X-ray Hit Map - Target Cu Method Spectrum Chip C10
Rate Rate 361.90 kHz/cm²
NTrigs N Trig 2956043 Trigger
NHits N Hits 16690 Hits
X-ray Hit Map - Target Zn Method Spectrum Chip C10
Rate Rate 407.20 kHz/cm²
NTrigs N Trig 2956195 Trigger
NHits N Hits 18779 Hits
X-ray Hit Map - Target Br Method Spectrum Chip C10
Rate Rate 618.90 kHz/cm²
NTrigs N Trig 2956196 Trigger
NHits N Hits 28542 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C10
Rate Rate 573.50 kHz/cm²
NTrigs N Trig 2956455 Trigger
NHits N Hits 26452 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C10
Rate Rate 293.40 kHz/cm²
NTrigs N Trig 2956176 Trigger
NHits N Hits 13529 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C10
Rate Rate 185.70 kHz/cm²
NTrigs N Trig 2956336 Trigger
NHits N Hits 8564 Hits