Test Date: 2015-04-08 11:04
Chip 14
X-ray Calibration - Target Cu Method Spectrum Chip C14
Center |
Center of Peak |
42.92 ± 0.15 |
Vcal |
TargetEnergy |
Energy of target Cu |
8047.82 |
eV |
TargetNElectrons |
Energy of target Cu |
2235.51 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
11.37 |
|
Rate |
Rate |
246.70 |
kHz/cm² |
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center |
Center of Peak |
46.57 ± 0.13 |
Vcal |
TargetEnergy |
Energy of target Zn |
8638.91 |
eV |
TargetNElectrons |
Energy of target Zn |
2399.7 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
12.5 |
|
Rate |
Rate |
292.10 |
kHz/cm² |
X-ray Calibration - Target Br Method Spectrum Chip C14
Center |
Center of Peak |
63.83 ± 0.11 |
Vcal |
TargetEnergy |
Energy of target Br |
11924.36 |
eV |
TargetNElectrons |
Energy of target Br |
3312.32 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
9.49 |
|
Rate |
Rate |
417.70 |
kHz/cm² |
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center |
Center of Peak |
95.89 ± 0.09 |
Vcal |
TargetEnergy |
Energy of target Mo |
17479.37 |
eV |
TargetNElectrons |
Energy of target Mo |
4855.38 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
7.71 |
|
Rate |
Rate |
454.60 |
kHz/cm² |
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center |
Center of Peak |
122.35 ± 0.13 |
Vcal |
TargetEnergy |
Energy of target Ag |
22162.92 |
eV |
TargetNElectrons |
Energy of target Ag |
6156.37 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
5.02 |
|
Rate |
Rate |
262.80 |
kHz/cm² |
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center |
Center of Peak |
139.81 ± 0.17 |
Vcal |
TargetEnergy |
Energy of target Sn |
25271.36 |
eV |
TargetNElectrons |
Energy of target Sn |
7019.82 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
4.07 |
|
Rate |
Rate |
170.20 |
kHz/cm² |
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope |
Slope |
49.248 ± 0.194 |
nElectrons/VCal |
Offset |
Offset |
132.665 ± 21.632 |
nElectrons |
X-ray Hit Map - Target Cu Method Spectrum Chip C14
Rate |
Rate |
246.70 |
kHz/cm² |
NTrigs |
N Trig |
2956043 |
Trigger |
NHits |
N Hits |
11376 |
Hits |
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate |
Rate |
292.10 |
kHz/cm² |
NTrigs |
N Trig |
2956195 |
Trigger |
NHits |
N Hits |
13472 |
Hits |
X-ray Hit Map - Target Br Method Spectrum Chip C14
Rate |
Rate |
417.70 |
kHz/cm² |
NTrigs |
N Trig |
2956196 |
Trigger |
NHits |
N Hits |
19261 |
Hits |
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate |
Rate |
454.60 |
kHz/cm² |
NTrigs |
N Trig |
2956455 |
Trigger |
NHits |
N Hits |
20966 |
Hits |
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate |
Rate |
262.80 |
kHz/cm² |
NTrigs |
N Trig |
2956176 |
Trigger |
NHits |
N Hits |
12121 |
Hits |
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate |
Rate |
170.20 |
kHz/cm² |
NTrigs |
N Trig |
2956336 |
Trigger |
NHits |
N Hits |
7848 |
Hits |