Test Date: 2015-04-08 11:04
Chip 15
X-ray Calibration - Target Cu Method Spectrum Chip C15
Center |
Center of Peak |
48.4 ± 0.13 |
Vcal |
TargetEnergy |
Energy of target Cu |
8047.82 |
eV |
TargetNElectrons |
Energy of target Cu |
2235.51 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
4.76 |
|
Rate |
Rate |
273.80 |
kHz/cm² |
X-ray Calibration - Target Zn Method Spectrum Chip C15
Center |
Center of Peak |
50.69 ± 0.12 |
Vcal |
TargetEnergy |
Energy of target Zn |
8638.91 |
eV |
TargetNElectrons |
Energy of target Zn |
2399.7 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
4.87 |
|
Rate |
Rate |
315.80 |
kHz/cm² |
X-ray Calibration - Target Br Method Spectrum Chip C15
Center |
Center of Peak |
67.96 ± 0.19 |
Vcal |
TargetEnergy |
Energy of target Br |
11924.36 |
eV |
TargetNElectrons |
Energy of target Br |
3312.32 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
4.72 |
|
Rate |
Rate |
443.70 |
kHz/cm² |
X-ray Calibration - Target Mo Method Spectrum Chip C15
Center |
Center of Peak |
103.35 ± 0.12 |
Vcal |
TargetEnergy |
Energy of target Mo |
17479.37 |
eV |
TargetNElectrons |
Energy of target Mo |
4855.38 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
3.24 |
|
Rate |
Rate |
475.30 |
kHz/cm² |
X-ray Calibration - Target Ag Method Spectrum Chip C15
Center |
Center of Peak |
131.78 ± 0.16 |
Vcal |
TargetEnergy |
Energy of target Ag |
22162.92 |
eV |
TargetNElectrons |
Energy of target Ag |
6156.37 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
3.8 |
|
Rate |
Rate |
273.00 |
kHz/cm² |
X-ray Calibration - Target Sn Method Spectrum Chip C15
Center |
Center of Peak |
150.61 ± 0.18 |
Vcal |
TargetEnergy |
Energy of target Sn |
25271.36 |
eV |
TargetNElectrons |
Energy of target Sn |
7019.82 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
2.79 |
|
Rate |
Rate |
174.50 |
kHz/cm² |
X-ray Calibration - Vcal Calibration Method Spectrum Chip C15
Slope |
Slope |
45.791 ± 0.199 |
nElectrons/VCal |
Offset |
Offset |
122.751 ± 24.497 |
nElectrons |
X-ray Hit Map - Target Cu Method Spectrum Chip C15
Rate |
Rate |
273.80 |
kHz/cm² |
NTrigs |
N Trig |
2956043 |
Trigger |
NHits |
N Hits |
12628 |
Hits |
X-ray Hit Map - Target Zn Method Spectrum Chip C15
Rate |
Rate |
315.80 |
kHz/cm² |
NTrigs |
N Trig |
2956195 |
Trigger |
NHits |
N Hits |
14563 |
Hits |
X-ray Hit Map - Target Br Method Spectrum Chip C15
Rate |
Rate |
443.70 |
kHz/cm² |
NTrigs |
N Trig |
2956196 |
Trigger |
NHits |
N Hits |
20463 |
Hits |
X-ray Hit Map - Target Mo Method Spectrum Chip C15
Rate |
Rate |
475.30 |
kHz/cm² |
NTrigs |
N Trig |
2956455 |
Trigger |
NHits |
N Hits |
21920 |
Hits |
X-ray Hit Map - Target Ag Method Spectrum Chip C15
Rate |
Rate |
273.00 |
kHz/cm² |
NTrigs |
N Trig |
2956176 |
Trigger |
NHits |
N Hits |
12588 |
Hits |
X-ray Hit Map - Target Sn Method Spectrum Chip C15
Rate |
Rate |
174.50 |
kHz/cm² |
NTrigs |
N Trig |
2956336 |
Trigger |
NHits |
N Hits |
8050 |
Hits |