Test Date: 2015-05-21 18:05
Chip 0
X-ray Calibration - Target Zn Method Spectrum Chip C0
Center Center of Peak 57.57 ± 0.08 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 14.3
Rate Rate 288.20 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C0
Center Center of Peak 115.35 ± 0.08 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 8.95
Rate Rate 449.30 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C0
Center Center of Peak 146.31 ± 0.1 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 7.65
Rate Rate 263.90 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C0
Center Center of Peak 165.63 ± 0.14 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 9.5
Rate Rate 169.90 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C0
Slope Slope 42.375 ± 0.059 nElectrons/VCal
Offset Offset -37.932 ± 6.162 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C0
Rate Rate 288.20 kHz/cm²
NTrigs N Trig 5910555 Trigger
NHits N Hits 26575 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C0
Rate Rate 449.30 kHz/cm²
NTrigs N Trig 5910983 Trigger
NHits N Hits 41433 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C0
Rate Rate 263.90 kHz/cm²
NTrigs N Trig 5941040 Trigger
NHits N Hits 24458 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C0
Rate Rate 169.90 kHz/cm²
NTrigs N Trig 5909762 Trigger
NHits N Hits 15659 Hits