Test Date: 2015-05-21 18:05
Chip 1
X-ray Calibration - Target Zn Method Spectrum Chip C1
Center Center of Peak 56.38 ± 0.09 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 10.62
Rate Rate 274.00 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C1
Center Center of Peak 112.64 ± 0.08 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 7.12
Rate Rate 446.80 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C1
Center Center of Peak 142.11 ± 0.11 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 8.63
Rate Rate 258.80 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C1
Center Center of Peak 161.18 ± 0.14 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 8.65
Rate Rate 171.90 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C1
Slope Slope 43.774 ± 0.069 nElectrons/VCal
Offset Offset -70.437 ± 7.249 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C1
Rate Rate 274.00 kHz/cm²
NTrigs N Trig 5910555 Trigger
NHits N Hits 25265 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C1
Rate Rate 446.80 kHz/cm²
NTrigs N Trig 5910983 Trigger
NHits N Hits 41203 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C1
Rate Rate 258.80 kHz/cm²
NTrigs N Trig 5941040 Trigger
NHits N Hits 23985 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C1
Rate Rate 171.90 kHz/cm²
NTrigs N Trig 5909762 Trigger
NHits N Hits 15849 Hits