Test Date: 2015-05-21 18:05
Chip 7
X-ray Calibration - Target Zn Method Spectrum Chip C7
Center Center of Peak 55.45 ± 0.08 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 15.75
Rate Rate 331.00 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C7
Center Center of Peak 112.75 ± 0.07 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 9.51
Rate Rate 528.10 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C7
Center Center of Peak 142.42 ± 0.09 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 8.65
Rate Rate 283.10 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C7
Center Center of Peak 161.69 ± 0.12 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 6.31
Rate Rate 179.20 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C7
Slope Slope 44.118 ± 0.114 nElectrons/VCal
Offset Offset -120.443 ± 15.055 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C7
Rate Rate 331.00 kHz/cm²
NTrigs N Trig 5910555 Trigger
NHits N Hits 30519 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C7
Rate Rate 528.10 kHz/cm²
NTrigs N Trig 5910983 Trigger
NHits N Hits 48699 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C7
Rate Rate 283.10 kHz/cm²
NTrigs N Trig 5941040 Trigger
NHits N Hits 26240 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C7
Rate Rate 179.20 kHz/cm²
NTrigs N Trig 5909762 Trigger
NHits N Hits 16520 Hits